Static weighing
The principle of the flow-rate measurement method by static weighing (for schematic diagrams of typical installations, see figures 1 A, 1 B, 1 C) is :
ü to determine the initial mass of the tank plus any residual liquid;
ü to divert the flow into the weighing tank (until it is considered to contain a sufficient quantity to attain the desired accuracy) by Operation of the diverter, which actuates a timer to measure the filling time;
ü to determine the final mass of the tank plus the liquid collected in it.
The flow-rate is then derived from the mass collected, the collection time and other data as discussed in clause 5 and annex A.
Figure IA - Diagram of an installation for calibration by weighing (static method, supply by a constant level head tank)
Figure 1C - Diagram of an installation for calibration by weighing
(static method, direct pumping supply)